Yearly Preventative Maintenance is Critical
A Yearly Preventative Maintenance Visit is critical to the proper operation and long term life of your SEAL Discrete Analyzer. During the annual PM Visit, the SEAL Factory Trained Service Engineer will replace the following items:
- Source Lamp
- All tubing including the waste tubing
- Aspiration Probe
- Sampler Probe
- Probe Washer
- Pump Tubing
- Syringe Body, Piston and O-ring
Clean and Lubricate (as appropriate):
- Cuvette
- Lenses in the Optics Module
- Wash Pump Roller Assembly
- Aspiration Pump Roller Assembly
- Aspiration Vertical and Horizontal Drives
- Vertical Sampler Drive
- Syringe Drive
- Aspiration Wash Bath
- Syringe Wash Bath
Electronic Checks / Adjustments:
- All Power Supplies are Checked
- Lamp Voltage Adjusted
- Detector Regulated Supplies
Mechanical Checks / Adjustments:
- Alignment of each motor position
- Sampler and Aspiration Probe Depth
- Aspiration Wash Pump Volume
- Aspiration Pump Volume
Instrument Validation Testing:
- Measuring Diagnostic - 30 repetition minimum
- Wavelength Linearity
- Wavelength Precision
- Optical Base Line